MDA Quantification in Jet Fuel

The MDAnalyzer™ is a portable analyzer able to provide real time determination of the presence and specific concentration of Metal Deactivator Additive (MDA) in an aviation turbine fuel sample at installation level or in a field environment outside of a laboratory. The MDAnalyzer enables a technician to determine the presence and level of MDA in a fuel sample in 20-30 minutes. The current process requires samples of suspected thermally unstable fuel or fuel previously doped with MDA to be collected on site and then shipped to a laboratory, which can take 2-10 days for results. During this transit and analysis time, fuel inventories and fuel servicing equipment are placed in a quality hold status, preventing the ability to support aircraft or ground vehicle generation and mission requirements.

MDAnalyzer-by-Axcend

Repeatability – Retention time repeatability is under 1% RSD, and peak area repeatability is under 4% RSD (10 ppm MDA in Jet Fuel with blank Heptane runs between samples).

MDA Spiked Into Jet A Fuel at 4ppm

jetfuelspike

  • 15 cm x 300 μm i.d., 3 μm Inertsil Diol
  •  A: heptane, B: ethyl acetate + 5% acetonitrile
  • 0.40 – 100% B in 9 min nonlinear program
  • 7.5 μL/min after 3 min equilibration
  • 315 nm

peakarea